Test range |
0.01-2*104Barrers |
Test flow |
0-100ml/min (200ml/min, 300ml/min Optional) |
Accuracy |
0.01Barrers |
Vacuum accuracy |
0.1Pa |
Test gas |
O2, N2, CO, CO2, H2, CH4, C2H4, C2H6, He and etc PS: DO NOT test oxygen together with flammable and explosive gas. Forbid to test corrosive gas. |
Pressure difference |
0-10bar (STD) |
System pressure resistance |
16bar |
Temperature range |
5℃~95℃, accuracy: ±0.1℃ (optional) |
Test area |
Φ20mm (3.14cm2) |
Sample thickness |
≤2mm |
Sample |
One piece |
Detector sensitivity |
≥5000mv.ml/mg or 10000mv.ml/mg |
Detector drift |
≤0.10mv. Noise: ≤0.015mv |
Loading gas pressure |
2bar |
Loading gas |
N2, H2, He, Ar and etc. |
Dimension |
320 mm(W)×625mm(D)×335mm(H) |
GC dimension |
590 mm(L)×480mm(W)×500mm(H) (STD) |
Power supply |
AC 220V 50Hz |
Delivery: Main unit, professional software,vacuum pump, tank regulator,
PCIE card, gas chromatograph, communication cable, vacuum
grease, gas pipeline.
Optional: Temperature controller, computer, DSM system.
Note: Specifications are subject to change without prior notice.